欢迎来Nano电子商城!
您好 [请登录]   [免费注册]
x
全部商品分类

BGI GK2.05旋风切割器手册

下载次数:398 浏览次数:680发布时间:2017-02-20 12:13
允许查看的会员等级:非会员   普通会员   高级会员   VIP会员   

文档详情

This cyclone is intended to be used with the BGI AFC-123 R/T or BGI 400 personal sampling pump or any

personal sampling pump meeting the following pressure volume characteristics:

GK2.05 (KTL)

37mm Media Type Flow Rate - LPM Pressure Drop cm of H2O

GF/A Fiberglass 4.0 10.2

0.8ìm MCE 4.0 33.0

Gelman Teflo, 2ìm 4.0 7.0

GK2.05 SH (KTL)

37mm Media Type Flow Rate - LPM Pressure Drop cm of H2O

GF/A Fiberglass 3.5 8.9

0.8ìm MCE 3.5 29.0

Gelman Teflo, 2ìm 3.5 6.2

At the stipulated flow rate of 4.0 lpm (GK2.05) or 3.5 lpm (GK2.05 SH) the instrument conforms to US EPA

(PM) 2.5 Standard with a 50% cut point of 2.5ìm.

Operation:

Remove the front portion of a 3 piece cassette (blue plug) by inserting a coin in the annular slot closest to the

blue plug and twisting. Save the portion removed to reinstall for filter protection after sampling. Press the

portion of the cassette containing the filter onto the tapered top of the cyclone, fitted with an orange “O” ring.

Next, remove the red plug from the back of the cassette and push in the taper luer plug, which is attached to

the hose.

The cyclone is clipped to the worker’s lapel and the long hose is run over the worker’s shoulder to the

personal sampler appropriately situated on the worker’s waist belt.

Proceed with sampling according to your experience, regulations, or the instructions for the personal sampler

being used. The procedures are the same as for a routine cassette dust sample, except that only the (PM)2.5

fraction actually reaches the filter surface. Selection of filter materials varies with different countries and

regulations.

Calibration GK2.05:

Attach your calibration device (rotameter or preferably a soap film meter or BGI TetraCal (formally triCal)) to

the inlet tube of the cyclone with a suitable section of 1/4 inch I.D. tubing. The cyclone should be set up for

normal use with a fresh filter. Turn the personal sampler on and adjust the flow rate to 4 liters per minute.

Because of the inlet tube used on this cyclone design, the old fashioned calibration jar is no longer necessary.

中国总代理:Nano电子商城

Tel: 4006609565

E-mail:19mro@19mro.com

http://www.19mro.com

Calibration GK2.05 SH

Press the calibration adapter (GK2.05CC) onto the body of the cyclone. Attach your calibration device

(rotameter or preferably The Challenger) to the calibration adapter of the cyclone with a suitable section of ¼

inch I.D. tubing. The cyclone should be set up for normal use with a fresh filter. Turn the personal sampler on

and adjust the flow rate to 3.5 liters per minute.

Cleaning:

After disconnecting the suction tubing, carefully remove the cassette and replace the cover and red plug.

Remove the grit pot by unscrewing it. Dispose of its contents and clean and dry it. Unscrew the top of the

cyclone. The parts can now be cleaned, thoroughly. Protracted soaking in soap/caustic solutions will

damage the aluminum components!

WARNING: Because of the almost infinite variety of dusts, which may be sampled with this device, it s not

possible to give specific recommendations for cleaning substances. Also, it must be noted to be careful not to

re-aerosolize hazardous materials when using compressed air for cleaning. Utilize good hygiene practices at

all times.

我们营业的时间9:00-18:00

  • 技术


关闭在线客服

Nano电子商城微信公众号手机商城 Nano电子商城微信公众号二维码

手机商城           微信商城

任何购物问题请联系Nano (Hongkong) Co.,Ltd  | 江苏省苏州市昆山开发区章基路189号科创基地2号厂房 Tel:021-52069907 email: 19mro@19mro.com

工作时间:周一至周五 8:00-18:00; 网站备案号 :沪ICP备12049383号; Copyright©2009-2012 19mro.com 版权所有